1. Enter Administrative mode, press and hold simultaneously.
2. Press selection buttons up or down to set pressure level from preset menu. For pressure level outside preset values, refer to the instructions for use - chapter "Change pressure setup mode".
3. Press "OK" to confirm and enter main display.
1. Press “Standby” for 3 seconds.
2. Clamp canister and dressing tubing.
3. Release and remove canister (if necessary, replace canister tubing).
4. Click new Invia Liberty canister into place.
5. Unclamp canister and dressing tubing
7. Seal used canister with a cap.
8. Dispose used canister in accordance with local procedures.
1. Administrative mode
2. Choose settings
1. Enter settings menu in Administrative mode, press and hold simultaneously.
2. Use selection buttons up or down to navigate.
3. Press "OK" to confirm setting.
1. Enter Administrative mode, press and hold simultaneously.
2. Press “Standby” for 3 seconds to put pump in standby mode.
3. Press “Change to Intermittent”. Default times for intermittent mode are 5 minutes on and 2 minutes off.